Gastat Pro Portable blood gas analyzer
Code: 14590
Manufacturer Ref: Gastat pro
7.192,00 €
(5.800,00 € + VAT)
Description:
By Japanese medical company, TechnoMedica
ACCESSORIES
Code:
14587
Price without VAT
118,00 €
Price with tax
146,32 €
Temporarily unavailable
Code:
14588
Price without VAT
131,00 €
Price with tax
162,44 €
Temporarily unavailable
Code:
14589
Price without VAT
121,00 €
Price with tax
150,04 €
Expected
Total:
7.192,00 €
Availability:
Temporarily unavailable
Delivery time:
Delivery 4 to 10 days
- Description
Description
Portable Type Blood Gas System with a variety of features.
Premium Quality by Japanese medical company TechnoMedica.
Compact & Lightweight.
Uses the newly developed Sensor Cards with built-in sensors and calibration fluid, ensuring high precision.
No danger of mix-ups in which results are applied to the wrong patient.
Blood gas and electrolytes test results within 45 seconds. (Measured time after sample injection)
Long Shelf-life Sensor Card: 2 months in room temperature (9-30 oC) / 12 months in refrigerator (2-8 oC)
Automatic Sample Transport System enables a sample to be transported by device driven and it achieves stability of test results.
Usage in operating rooms, emergency rooms, clinics, etc.
Its compact and lightweight design, can make it to be used, even inside the ambulance. Immediate use, anytime.
The calibration fluid in the Sensor Card activates automatic calibration with every measurement, allowing immediate use without preparation, anytime.
Specifications:
Sample Type:
- Human blood (whole blood, serum, plasma)
- Dialysate
- ISE-CRS
- GASTROL-QC
Measured parameters:
- pH : 6.000 - 8.000
- pCO2 : 5.0 - 250.0 Torr
- pO2 : 5.0 - 720.0 Torr
- cNa+ : 50.0 - 200.0 mmol/L
- cK+ : 1.0 - 12.0 mmol/L
- cCl- : 50.0 - 200.0 mmol/L
- cCa2+ : 0.25 - 5.0 mmol/L
- Hct : 10.0 - 75.0%
Calculated parameters: cHb, cHCO3-std, cBE(ecf), cBE(B), cBB, sO2(est), ctCO2(P), ctO2(a), cCa2+(7.4), pO2(A-a), RI,AG, pH(T), pCO2(T), pO2(T)
Sample volume: Over 90 µL
Analysis time: 135 seconds (90 seconds calibration + 45 seconds analysis)
Calibration: 1 point calibration before each sample analysis
Display: Color touch LCD
Printer: 20 digits thermal (58 mm)
Data capacity: 5000 results
Interface: USB
Sensor card shelf life: 2 months in room temperature (9-30 degrees Celsius) / 12 months in refrigerator (2-8 degrees Celsius)
Operating environment Temperature: 10-32 degrees Celsius
Humidity: 20 - 80 % (No condensation)
Dimensions: 100(W) - 220(D) - 90(H) mm
Sensor card: 49.8(W) - 77.0(D) - 8.2(H) mm
Weight: 1 kg
Power:
- AA batteries 4 primary batteries/AA rechargeable battery 4 secondary batteries
- AC 100 - 240V 50/60Hz 0.58A (AC adapter)